ULTRAPEN PT3 ORP/REDOX & Temperature Pen
Range: | -1000 mV to +1000 mV | 0 – 71°C / 32 – 160°F |
Accuracy: | 10 mV | ± 0.1ºC / ± 0.1ºF |
Resolution: | 1 mV ORP | 0.1ºC / 0.1ºF |
Time to Reading Stabilization: 10 – 45 seconds | ||
Power Consumption: Active Mode 37 mA, Sleep Mode 2 µA | ||
Temperature Compensation: Automatic In Calibration Mode From 15ºC to 30ºC* | ||
Physical Dimensions: 17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D | ||
Weight: 50.4 g / 1.78 oz. (without soaker cap and lanyard) | ||
Case Material: Anodized Aircraft Aluminum with Protective Coating | ||
Battery Type: N type, Alkaline, 1.5V | ||
Calibration Standard Solutions: ORP80, ORP260, ORP470 | ||
Operating/Storage Temperature: 0 – 55ºC or 32 – 131ºF | ||
Enclosure Ratings: IP67 and NEMA 6 | ||
EN61236-1: 2006 – Annex A: 2008: Electrostatic discharge to the PT3 may cause it to spontaneously turn on. If this occurs, the PT3 will turn off. |
* Temperature compensation in calibration mode: Temperature affects the reaction potentials for all chemicals differently. True ORP is the direct measurement of electron activity during an oxidation-reduction reaction, regardless of temperature. However, for maximum accuracy and ease of calibration, the Myron L Company has developed three calibration standard solutions with known dissolved species. We derived the temperature compensation (from 15C to 30C) for those solutions, and embedded automatic temperature compensation into the calibration function of your PT3. Note: To verify calibration while in measurement mode, you must manually correct for any variation in temperature. Example: @25C, ORP2602OZ calibration solution will read 260mV, however @ 20.0C ORP2602OZ will read 265mV.